A validation study.

Checkout our iOS App for a better way to browser and research.

§ 761.395 A validation study.

(a) Decontaminate the following prepared sample surfaces using the selected testing parameters and experimental conditions. Take a standard wipe sample of the decontaminated surface.

(1) At least one uncontaminated surface. The surface levels of PCBs on the uncontaminated surface must be <1 µg/100 cm2.

(2) At least seven contaminated surfaces.

(b)

(1) Use SW-846, Test Methods for Evaluating Solid Waste methods for sample extraction and chemical analysis as follows: Use Method 3500B/3540C or Method 3500B/3550B for the extraction and cleanup of the extract and Method 8082 for the chemical analysis, or methods validated under subpart Q of this part.

(2) Report all validation study surface sample concentrations on the basis of micrograms of PCBs per 100 cm2 of surface sampled.

(c) Following completion of the validation study, measurements from the contaminated surfaces must have an arithmetic mean of ≤10 µg/100 cm2. If the arithmetic mean is >10 µg/100 cm2, then the validation study failed and the solvent may not be used for decontamination under § 761.79(d)(4) according to the parameters tested.


Download our app to see the most-to-date content.